MRC EMF-831SD Bedienungsanleitung

SD card real time datalogger, RS232/USB
RF EMF Strength Meter
Model : EMF-831SD
Your purchase of this
Personal RF EMF
Strength Meter
with SD CARD
DATALOGGER marks a
step forward for you
into the field of
precision measurement.
Although this METER is
a complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.
OPERATION MANUAL
3, Hagavish st. Israel 58817 Tel: 972 35595252, Fax: 972 3 5594529
MRC. 2.20

TABLE OF CONTENTS
1. FEATURES..................................................................................................................................
1
2. SPECIFICATIONS........................................................................
...................................................
2
3. FRONT PANEL DESCRIPTION..................................................................
...........................
5
4. MEASURING PROCEDURE....................................................................
...........................
6
4-1 RF EMF Strength Function selection and measurement
..............................................................
6
4-2 Data Hold..........................................................
...........................................................
7
4-3 Data Record ( Max., Min. reading )...............................................................................................................
7
5. DATALOGGER................................................................................
......................................................
8
5-1 Preparation before execute datalogger function...............................................................................
8
5-2 Auto Datalogger ( Set sampling time ≧ 1 second ).......................
...........................
9
5-3 Manual Datalogger ( Set sampling time = 0 second )....................
...........................
10
5-4 Check time and sampling time information..........................................................................................................
10
5-5 SD Card Data structure..................................................................................................
11
6. Saving data from the SD card to the computer..................................................................
12
7. ADVANCED SETTING...............................................................................
...........................
14
7-1 Set clock time ( Year/Month/Date, Hour/Minute/ Second )........................................
14
7-2 Set sampling time ................................................................................................................
15
7-3 Auto power OFF management ..........................................................
...........................
15
7-4 Set beeper Sound ON/OFF.................................................................
...........................
15
7-5 Decimal point of SD card setting................................................................................................
16
7-6
SD memory card format....................................................................................................
16
7-7 Set mesurement value gain .................................................
...........................
17
7-8 Set the alarm value .................................................................................
..........
17
7-9 Select the alarm function yes or no .................................................
...........................
17
8. POWER SUPPLY from DC ADAPTER.....................................................
...........................
18
9. BATTERY REPLACEMENT..............................................................................................
18
10. RS232 PC serial interface.................................................................................
...........................
18
11. PATENT.................................................................................................................................................
20

1. FEATURES
* Measurements range : 50MHz to 3.5GHz.
* For electromagnetic field strength measurement including mobile
phone base station antenna radiation, RF power measurement for
transmitters, wireless LAN (Wi-Fi) detection/installation, wireless
communication applications (CW, TDMA, GSM, DECT) and
microwave leakage
* The meter is a broadband device for monitoring high-frequency radiation
in the specific frequency ranges of 900MHz, 1800MHz, and 2.7GHz.
Other measurements can be made,using the entire range of 50MHz to 3.5GHz.
* Non-directional (isotropic) measurement with three-channel
(triaxial) measurement probe
* Microcomputer circuit provides intelligent function and high accuracy.
* Real time SD memory card Datalogger, it Built-in Clock
and Calendar, real time data recorder, sampling time set
from 1 second to 3600 seconds.
* Manual datalogger is available ( set the sampling
time to 0 second ), during execute the manual datalogger
function, it can set the different position ( location ) No.
( position 1 to position 99 ).
* Innovation and easy operation, computer is not need to
setup extra software, after execute datalogger, just take
away the SD card from the meter and plug in the SD card
into the computer, it can down load the all the measured
value with the time information ( year/month/date/
hour/minute/second ) to the Excel directly, then user can
make the further data or graphic analysis by themselves.
* SD card capacity : 1 GB to 16 GB.
* Can default auto power off or manual power off.
* Data hold, record max. and min. reading.
* Microcomputer circuit, high accuracy.
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter.
* RS232/USB PC COMPUTER interface.
* Heavy duty & compact housing case.
1

2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 30 mm
Measurement Digital, triaxial measurement.
method
Sensor type Electrical-field sensors
resolution 0.1mV/m, 0.1uA/m, 0.1uW/m^2, 0.001uW/cm^2
Range autorange
Datalogger Auto 2 to 3600 seconds
Sampling Time
Setting range @ Sampling time can set to 1 second,
but memory data may loss.
Manual Push the data logger button
once will save data one time.
@ Set the sampling time to
0 second.
@ Manual mode, can also select the
1 to 99 position ( Location ) no.
Data error no. ≦ 0.1 % no. of total saved data typically.
Memory Card SD memory card. 1 GB to 16 GB.
Advanced * Set clock time ( Year/Month/Date,
setting Hour/Minute/ Second )
* Set sampling time
* Auto power OFF management
* Set beep Sound ON/OFF
* Decimal point of SD card setting
* SD memory card Format
*Set temperature unit to ℃ or ℉
2

Over Indication Show " - - - - ".
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
Sampling Time Approx. 1 second.
of Display
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Power off Auto shut off saves battery life or
manual off by push button.
Operating -20 ~ 50 ℃
Temperature
Operating Less than 85% R.H.
Humidity
Power Supply * Batteries( UM3, AA ) x 6 PCs, or equivalent.
* AC/DC power adapter .
(adapter is optional ).
Power Current Normal operation ( w/o SD card save data) :
Approx. DC 24 mA.
When SD card save the data :
Approx. DC 50 mA.
Weight 244 g/0.54 LB ( meter only ).
Dimension 177 x 68 x 45 mm
(7.0 x 2.7x 1.8 inch)
Accessories *
Instruction manual......................................
1 PCS
Included
Optional * SD Card ( 4 GB )
Accessories * USB cable, USB-01.
* RS232 cable, UPCB-02.
* Data Acquisition software,
* SW-U801-WIN., SW-E802.
AC to DC 9V adapter.
* Hard carrying case, CA-06.
* Soft carrying case, CA-05A.
3

2-2 Electrical Specifications (23±5
℃
, 25% ~ 75 % RH)
Frequency range: 50 MHz to 3.5GHz
Specified measurement range: 20 mV/m to 25.0 V/m
CW signal (f > 50MHz): 53 uA/m2 to 66.3mA/m
1 uW/m2 to 7.15W/m2
0 uW/cm2 to 0.715mW/cm2
Units mV/m, V/m, µA/m, mA/m, µW/m2,
mW/m2, W/m2, µW/cm2, mW/cm2
Display resolution: 0.1mV/m, 0.1uA/m, 0.1uW/m^2,
0.001uW/cm^2 , 0.001W/cm2
Dynamic range Typically 75dB
Absolute error 1.0dB
at 1 V/m & 2.45GHz
Frequency response 1.0dB (1.9GHz to 3.5GHz)
Sensor (with typical CAL factors) 2.4dB (50MHz to 1.9GHz)
Isotropy deviation Typically 1.0dB (f>2.45GHz)
Overload limit 0.42 mW/cm^2 per axis.
Thermal response (0 to 50°) ± 0.5dB
4

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display. 3-15 Battery compartment/Cover
3-2 Power Button 3-16 Tripod Fix Nut
3-3 Hold Button 3-17 RF EMF detect seneor
3-4 REC Button
3-5 ▲ Button ( UNIT Button)
3-6 ▼ Button ( X,Y,Z,XYZ Function Button )
3-7 Time Button (SET Button )
3-8 Enter Button ( Logger Button)
3-9 SD card socket
3-10 RS-232 output terminal
3-11 Reset Button
3-12 DC 9V adapter socket
3-13 Stand
3-14 Battery Cover Screws
5

4. MEASURING PROCEDURE
E-field sensors
The 3-channel sensor is located at the top of the meter. The three voltages
generated by the sensor are fed back to the meter.
The measurement of the field is done by moving the aerial of the sensor in
the desired measured environment. A direct wide band
measurement is obtained of the field that the measurement sensor is
subjected to. To find the value of the field emitted by a source of interference,
simply point the aerial towards it and get as close as possible (the value of the
field is inversely proportional to the distance of the sensor/emission source).
The operator must take care not to be between the source of disturbance and
the zone to be checked. The human body shields electromagnetic fields.
4-1 RF EMF Strength Function selection and measurement
1) Turn on the meter by pressing the " Power Button "
( 3-2, Fig. 1 ) > 1.5 seconds continuously.
*Pressing the " Power Button " ( 3-2, Fig. 1 )
continuously and > 1.5 seconds again will turn off the
meter.
2) measurement Function select
pressing and hold the " Function Button "( 3-6, Fig. 1 ),
to select X axis,Y axis, Z axis, and XYZ axis measurement Function.
3) measurement UNIT select
pressing and hold the " UNIT Button "( 3-5, Fig. 1 ),
to select …
1mV/m, When Test values > 999.9 ,the unit auto chang to V/m unit ,
2µA/m, When Test values > 999.9 ,the unit auto chang to mA/m unit ,
3µW/m^2, When Test values > 999.9 ,the unit auto chang to mW/m^2 unit ,
mW/m^2, When Test values > 999.9 ,the unit auto chang to W/m^2 unit ,
4
µW/cm^2, When Test values > 999.9 ,the unit auto chang to mW/cm^2 unit ,
4) measurement Procedure
1. Hold the meter at arm’s length.
2. Make several measurements at various locations around the work place
or the interested areas as described above. This is particularly important
if the field conditions are unknown.
6

3. Pay special attention to measuring the neighboring vicinity for possible
radiation sources. Apart from active sources, those components connected
to a source may also act as radiators. For example, the cables used in
diathermy equipment may also radiate electromagnetic energy. Note
that metallic objects within the field may locally concentrate or amplify the
field from a distant source.
4-2 Data Hold
During the measurement, press the " Hold Button " ( 3-3,
Fig. 1 ) once will hold the measured value & the " HOLD " symbole
will be light, Press the " Hold Button " once again will release the data
hold function , ( 3-20, Fig. 1) HOLD symbole will be quenched
4-3 Data Record ( Max., Min. reading )
1) The data record function records the maximum and minimum readings.
Press the " REC Button " ( 3-4, Fig.1 ) once to start the Data Record
function and there will be a " REC " symbol indicator "
will be light.
2) With the " REC " Symbol is light on the display :
a) Press the " REC Button " ( 3-4, Fig. 1 ) once, the "
REC and MAX Symbol indicator will be light ,
and the maximum value will appear on the display.
b) Press the " REC Button " ( 3-4, Fig. 1 ) again, the
" REC and MIN " Symbol indicator will be light ,
and the minimum value will appear on the display.
c) To exit the memory record function, just press the
" REC " button for 1.5 seconds at least. The display will
revert to the current reading, red led indicator will be quenched
7

5. DATALOGGER
5-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 G to 16 G, optional ),
insert the SD card into the " SD card socket " ( 3-9, Fig. 1).
The front panel of the SD card should face against the
the down case.
*It recommend use memory card
≦
4 GB.
b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first. ,
please refer chapter 7-6, (page 16 ) .
*
It recommend strongly, do not use memory cards that
have been formatted by other meter or by other installation
(such as camera…) .
Reformat the memory card with your meter.
*
If the SD memory card exist the trouble during
format by the meter use the computer to
reformat again can fix the problem .
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-1 ( page 14 ).
d. Decimal for the numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 16.
8
Andere MRC Messgerät Handbücher

MRC
MRC HE218B-EX Bedienungsanleitung

MRC
MRC DW-6163 Bedienungsanleitung

MRC
MRC MO-2001 Bedienungsanleitung

MRC
MRC SPECTRO-96 Bedienungsanleitung

MRC
MRC LX-1102 Bedienungsanleitung

MRC
MRC AM-4233SD Bedienungsanleitung

MRC
MRC WA-2017SD Bedienungsanleitung

MRC
MRC PAM-9212SD Bedienungsanleitung

MRC
MRC PMG-302 Bedienungsanleitung

MRC
MRC NIR-CS-3000BT Bedienungsanleitung

MRC
MRC DO-5510 Bedienungsanleitung

MRC
MRC CD-4307SD Bedienungsanleitung

MRC
MRC PM-9100 Bedienungsanleitung

MRC
MRC PVB-820 Bedienungsanleitung

MRC
MRC RSW-923 Bedienungsanleitung

MRC
MRC LX-1148SD Bedienungsanleitung

MRC
MRC CLRM-110 Bedienungsanleitung

MRC
MRC FG-6020SD Bedienungsanleitung

MRC
MRC UT-FLU5B Bedienungsanleitung

MRC
MRC COH-9905SD Bedienungsanleitung