1
9
PUMP FEATURES
1 Pump Features
Guided handling sequences (wizards)
Pump therapy sometimes involves complex
steps. For several functions which need to
be performed in a specific order, your pump
offers pre-programmed wizards to guide
you through the process.
You will find wizards to help you with:
ºInserting a new cartridge
ºFilling the infusion set
ºInserting a new battery
ºPairing
ºBluetooth®wireless technology
3 basic modes
RUN mode: When working normally, your
pump is in RUN mode. In this mode your
pump is delivering insulin to your body in
terms of the basal rate, boluses and
Temporary Basal Rates (TBR).
PAUSE mode: When you remove the battery
or cartridge while your pump is in RUN
mode, or after some maintenance
messages, your pump switches to PAUSE
mode. The insulin delivery stops
automatically and restarts as soon as you
confirm the respective messages and insert
a new battery or cartridge where required.
Any current Temporary Basal Rates and
boluses will continue. Furthermore, the
basal rate missed while the pump was in
PAUSE mode will be delivered on top of the
ongoing basal rate once the pump is back in
RUN mode.
STOP mode: When the pump is in PAUSE
mode for more than 15 minutes, it switches
to STOP mode. The Pump stopped screen
appears and the pump stops completely and
cancels all current Temporary Basal Rates
and boluses. From STOP mode, you have to
restart your pump from the main Menu and
program your Temporary Basal Rates and
boluses again. See chapter 3.2, page 34.
If you want to stop your pump for any
reason, you can also switch it to STOP mode
using the main Menu.
Adjust the display to suit your needs
You can adjust the brightness of your
pump’s full colour display to best suit your
needs. You can also change its background
colour to your personal preference. Rotating
the display allows you to read it from an
awkward angle (for example, when the
pump is attached to your belt). See
chapter 8.3.5, page 74.
Enlarge the display texts for better
readability
Using the zoom key you can enlarge the size
of the display text for better readability. See
chapter 1.1.4, page 15.
User profiles for special situations
With the user profile Expressive enabled, the
scrolling functions of the pump are disabled
so that all values can only be programmed
in distinct increments. Additionally, sound
patterns will help you verify that you
programmed a value correctly. This may be
helpful if you have impaired eyesight. For
man_06719414001_K_en-UK.indb 9 27.06.2019 22:16:20