Evident SZ61 Bedienungsanleitung

INSTRUCTIONS
This instruction manual is for the Olympus Zoom Stereo Microscopes. To ensure the safety,
obtain optimum performance and to familiarize yourself fully with the use of this microscope, we
recommend that you study this manual thoroughly before operating the microscope.
Retain this instruction manual in an easily accessible place near the work desk for future reference.
For details of products included in the configuration of this system, see “1 MODULES AND
NOMENCLATURE” (P.3) and “8 OPTIONAL MODULES” (P. 18).
ZOOM STEREO MICROSCOPES
SZ61/SZ61-60
SZ61TR
SZ51/SZ51-60
Optical Microscope and Accessory


SZ2 series
IMPORTANT....................................................................................................................................................... 1
1 MODULES AND NOMENCLATURE .......................................................................................... 3
2 SUMMARY OF OBSERVATION PROCEDURE.................................................................. 5
2-1 Preparation.............................................................................................................................................................. 5
2-2 Observation Procedure................................................................................................................................ 5
3 OPERATION.................................................................................................................................................. 6
3-1 Base.............................................................................................................................................................................. 6
1Using the Stage Plate................................................................................................................................................................6
2Adjusting the Focus Adjustment Knob Tension...............................................................................................................................6
3-2 Microscope Body.............................................................................................................................................. 6
1Adjusting the Interpupillary Distance...........................................................................................................................6
2Adjusting the Diopter (Zoom Confocality Adjustment)................................................................................7
3Using the High/Low Zoom Magnification Stopper .........................................................................................8
4Using the Eye Shades ..............................................................................................................................................................8
5Mounting the Eyepiece Micrometer Disk.................................................................................................................9
6Using the Auxiliary Objectives........................................................................................................................................10
7Selecting the Light Path (SZ61TR) ..............................................................................................................................10
8Adjusting the Confocality of the Camera (SZ61TR)..................................................................................... 11
4 TROUBLESHOOTING GUIDE ......................................................................................................12
5 SPECIFICATIONS ...................................................................................................................................13
6 OPTICAL PERFORMANCE.............................................................................................................14
7 ASSEMBLY...................................................................................................................................................15
7-1 Assembly Diagram .........................................................................................................................................15
7-2 Detailed Assembly Procedures...........................................................................................................16
8 OPTIONAL MODULES .......................................................................................................................18
8-1 BX Stage Adapter Type 1 SZX-STAD1 ..........................................................................................18
8-2 Stage Adapter Type 1 SZH-STAD1..................................................................................................20
8-3 Gliding Stage SZH-SG................................................................................................................................. 21
8-4 Cup Stage SZH-SC.......................................................................................................................................22
CONTENTS


1
SZ2 series
IMPORTANT
1. After the equipment has been used in an observation of a
specimen that is accompanied with a potential of infection, clean
the parts coming in contact with the specimen to prevent infection.
· Moving this microscope is accompanied with the risk of
dropping the specimen. Be sure to remove the specimen
before moving this product.
The SZ2 series of stereo microscopes are provided with ESD (Electro-Static Discharge) protection specifications. They are
provided with electroconductive coating in their external finish to reduce the surface resistance and grounding lead wires
in their standard base or frame to eliminate static electricity from them.
To maintain the ESD protection performance, always use the microscope in combination with the
modules described in this manual or options having the ESD protection specifications. Otherwise, the
grounding will not work properly.
·The eye shades of the eyepieces are not provided with the ESD protection performance. Be sure to
remove them
·The SZ2-SPBW stage plate is provided with the ESD protection performance only in the black surface.
NOTE
SAFETY PRECAUTIONS
b
a
· Hold the bottom of the stand with one hand while holding base column with the other hand to avoid tilting the
microscope.
· In case the specimen is damaged by erroneous operation, promptly take the infection prevention measures.
2. When the optional 0.5X auxiliary objective ais used, the microscope become unstable due to the use of the SZ2-
ET auxiliary sleeve b. Be careful not to topple down the microscope. )
In addition, when a camera is mounted on the SZ61TR, special care is required because the microscope becomes
more unstable.
1Getting Ready
1. A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight, high temperature and humidity, dust or
vibrations. (For the operating conditions, see “SPECIFICATIONS” on page 13.)
3. Be careful not to leave stains and fingerprints on the lens surfaces.
Contamination of the lens and mirror deteriorates the view of images.
4. Do not turn the left and right zoom adjustment knobs in the opposite directions, as this will result in a failure.
5. The rubber parts of the eyepiece sleeves are fragile and should be handled carefully. If they are damaged, dirt will
penetrate inside the microscope.

2
2Maintenance and Storage
1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower
and wipe gently using a piece of cleaning paper (or clean gauze).
If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available
absolute alcohol.
Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks –– for
example, electrical equipment that is being switched on or off.
Also remember to always use it only in a well-ventilated room.
2. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To
clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.
4. When disposing of the microscope. Check the regulations and rules of your local government and be sure to
observe them.
CAUTION
If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition,
the equipment may also be damaged. Always use the equipment as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
CAUTION : Indicates that failure to follow the instructions in the warning could result in
bodily harm to the user and/or damage to equipment (including objects in the
vicinity of the equipment).
NOTE
: Indicates that failure to follow the instructions could result in damage to equipment.
TIP
: Indicates commentary (for ease of operation and maintenance).
3Caution
4Intended use
This product has been designed to be used to observe magnified images of specimens in various routine work and
research applications.
This includes the observation of living cells or of specimen taken from tissues to gain physiological or morphological
information at hospitals or laboratories.
Typical field of applications are genetics, human blood and tissue examination, neurology, pharmacology and cellular biology.
Further applications of this device include measurement and imaging for materials research, precision manufacturing, electronics
design and medical device fabrication. Additional industrial applications are added by individual companies and researchers.
Do not use this product for any purpose other than its intended use.
This product complies with the requirements of directive 98/79/EC concerning in vitro diagnostic medical
devices. CE marking means the conformity to the directive.
This product is applied with the requirements of EMC standard IEC/EN61326-2-6 and IEC/EN61326-1
concerning electromagnetic compatibility.
This product complies with the emission and immunity requirements described in IEC61326 series. The
electromagnetic environment should be evaluated prior to operation of this product.

3
SZ2 series
1 MODULES AND NOMENCLATURE
Eyepiece clamping knob
(x 2)
Accessory mount thread
(x 5)
Zoom adjustment knob (Page 7and 8)
· SZ61/61-60/TR: 0.67X to 4.5X
· SZ51/51-60: 0.8X to 4X
Focus adjustment knob
(Page 6)
Stroke : 120 mm
Eye shade (Page 8)
Auxiliary objective mount thread
Optional auxiliary objectives
· 110ALK0.3X
· 110ALK0.4X
· 110AL0.5X-2
· 110AL0.62X
· 110AL0.75X-2
· 110AL1.5X
· 110AL2X-2
Optional stage mounting holes
M4 screw (x 2)
Accessory mount holes
Eyepiece
· WHSZ10X
· WHSZ10X-H
· WHSZ15X-H
· WHSZ20X
· WHSZ20X-H
· WHSZ30X-H
Body clamping knob
(Threaded for accessory mounting)
Microscope Body
· SZ61
· SZ61-60
· SZ61TR
· SZ51
· SZ51-60
Stage Plate
· SZ2-SPBW
(ESD protection on black surface)
· SP-C
Olympus classifies SZ61, SZ61-60, SZ61TR, SZ51 and SZ51-60 as an optical microscope and other units as optical
microscope accessories.
Stand
Standard stand: SZ2-ST
LED illuminator stand: SZ2-ILST*
LED transmitted light illumination:
SZX2-ILLTQ*
SZX2-ILLTS*
A separate instruction manual is available for the module marked *.

4
Observation Tube of SZ61TR
C-mount (Page 11)
C-mount clamping screw
Confocality adjustment clamping
screw (Page 11)
(x 2)
Light path selector lever (Page 10)
Pulled out : Camera path 100%
(Right eyepiece sleeve
light path 0%)
Pushed in : Right eyepiece sleeve
light path 100%

5
SZ2 series
2SUMMARY OF OBSERVATION PROCEDURE
Ref. Page
1. Confirm the mounting, particularly that of the microscope body. .......................................................................................................................................(Page 15 — 17)
2. Confirm that the eyepieces are positioned correctly with respect to the standard base...................................................................................(Page 16)
3. Adjust the rotation tension of the focus adjustment knob............................................................................................................................................................................(Page 6)
4. Prepare the light source as required.
2-2 Observation Procedure
2-1 Preparation
Place the specimen on the stage plate. (Page 6)
Adjust the interpupillary distance. (Page 6)
1
2
Adjust the eyepiece diopter. (Page 7)
This operation is not possible with the eyepieces without
helicoids (WHSZ10X/20X).
The operation is slightly difference when a micrometer is mounted in
one of the eyepieces with helicoids.
3
NOTE
Set the zoom adjustment knob afor the lowest magnification and
rotate the focus adjustment knob bto bring the specimen into
approximate focus.
Rotate the zoom adjustment knob ato the target magnification
and rotate the focus adjustment knob bto bring the specimen into
accurate focus.
4
5
ab

6
3 OPERATION
3-1 Base
When reflected light illumination is used, the stage plate can be placed
either the white or black surface facing up depending on which side
makes the specimen easier to view.
·To maintain the ESD performance, use the black surface of
the SZ2-SPBW stage plate.
·When transmitted light illumination is used, use the optional
SP-C transparent glass stage plate.
NOTE
This operation is intended to facilitate the rotation of the
knobs while preventing spontaneous drop of the microscope
body. It is recommended to set the knob tension to a slightly
higher level than the point where spontaneous drop occurs.
Hold the left and right focus adjustment knobs awith both hands,
fix the left knob and rotate the right knob. The rotation tension of the
knobs increases or decreases according to the direction in which the
right knob is rotated.
TIP
1
1Using the Stage Plate
2Adjusting the Focus Adjustment Knob Tension
a
Clamp.
To increase
To decrease
3-2 Microscope Body
While holding the left and right eyepiece sleeves awith both hands,
look through the eyepieces and adjust for binocular vision until the
left and right fields of view coincide completely.
1Adjusting the Interpupillary Distance
a
Dieses Handbuch passt für folgende Modelle
4
Inhaltsverzeichnis
Andere Evident Mikroskop Handbücher
Beliebte Mikroskop Handbücher anderer Marken

Thermo Scientific
Thermo Scientific Nicolet iN10 Leitfaden

National
National DC3-163 Bedienungsanleitung

KERN Optics
KERN Optics OPE-1 Bedienungsanleitung

OPTIKA MICROSCOPES
OPTIKA MICROSCOPES B-150D Bedienungsanleitung

Prior
Prior Lux POL Bedienungsanleitung

Opitec
Opitec Smartscoop 115615 Bedienungsanleitung

VWR
VWR VisiScope 384 Series Bedienungsanleitung

Nikon
Nikon ECLIPSE E200 POL Bedienungsanleitung

Leica
Leica DI C800 Handbuch

ThermoFisher Scientific
ThermoFisher Scientific Continuµm Bedienungsanleitung

ThermoFisher Scientific
ThermoFisher Scientific Continuµm Bedienungsanleitung

Olympus
Olympus SZ61 Bedienungsanleitung