MR YD-200 Bedienungsanleitung

YD-200 Urine Analyzer
User Manual
Jilin Yida Medical Equipments Co.,Ltd

YD-200 Urine Analyzer User Manual
1
Content
Chapter 1. Introduction...................................................................... ..2
Chapter 2. Technical Specifications, structure, application range……4
Chapter 3. Installation and functions…………………………………6
Chapter 4. Set up……………………………………………………..10
Chapter 5. Testing options…………………………………………...19
Chapter 6. Worktable………………………………………………. . 20
Chapter 7. Test method……………………………………………….21
Chapter 8 .Record query……………………………………………...26
Chapter 9. Maintenance………………………………………………27
Chapter 10. Malfunction solution…………………………………… 28
Appendix A. Output value of YD-200 Urine Analyzer……………...29

YD-200 Urine Analyzer User Manual
2
Chapter 1 General Description
YD-200 urine analyzer is a kind of adopt latest optoelectronic technology and
microprocessor technology for Clinical urine test; it is user-friendly, and accurate.
URS-11A urine strip is dedicated for YD-200. YD-200 can provide 11 items results of
quantitative or semi-quantitative for Urobilinogen, Bilirubin, Ketone, Blood, Protein,
Micro-albumin ,Nitrite, Leucocytes, Glucose, Specific gravity, pH, Vitamin C.
Symbol information: The symbols can be seen on the analyzer and introduction as
follows:
Biohazard:Pay attention to the protection and treatment.
Shock hazard: Pay attention to the electric shock risk, avoid touching
dangerous charged items.
Danger:Attention Please Refer to their requirement of user manual.
Protective earthing: Protect earthing terminal, must ensure it cont act the
grounding.
IVD device
Intended use
It can provide a qualitative or semi-quantitative result of content among the personal
urine sample for medical institutions.
Instrument life
Normally, instrument life is 5 years at least, if the daily maintenance does well.
Instrument store
Keep in a cool, dry place.

YD-200 Urine Analyzer User Manual
3
Chapter 2 Technical Specifications, Structure and Application Range
2.1 Technical Specifications
Testing Principle:the method of high luminous cold light source reflectivity
Monochromatic color: white
Test items:
Glucose
Bilirubn
Ketone
Gravity
Vitamin C
pH
Protein
Urobilinogen
Nitrite
Blood
Leucocytes
GLU
BIL
KET
SG
VC
pH
PRO
URO
NIT
BLD
LEU
Test speed:
Fast mode:120 specimens/hour
Slow mode: 60 specimens/hour
Data Memory:1,000 patient results
Environment Requirements:Required temperature range is 0℃~40℃
Relative humidity ≤85%
Baud rate:1200,9600,19200,115200bps
Power Supply:AC220V/110V, 50/60Hz
Power:30VA
Net Weight:2Kg
Dimensions:L×W×H :285×237×171mm
2.2 Structure
DY-200 is made up by communication unit, control unit, power system, mechanical
unit, display unit, photoelectric scanning unit, keyboard, printer and some other parts.
2.3 Intended use
It can provide a qualitative or semi-quantitative result of content among the personal
urine sample for medical institutions.

YD-200 Urine Analyzer User Manual
4
It must take the dedicated urine strips. Urine strips are validity of consumables. Please
pay attention to the whether it is in the period of validity and color of zone is changed.
If so, the urine strips cannot be used. Urine strips is not a part of instrument. User are
required to read the User's Guide before taking use of the instrument, otherwise the
protection of the instrument itself could be broken, the accuracy of result could not be
ensured.
Chapter 3 Installation and Functions
3.1 Name of Part(Figure3-1 3-2)
Figure 3-1 Figure 3-2
①LCD Screen:Display reminder, results and so on.
②Keyboard:Realize man-machine dialogue.
③Printer: Micro, low noise, high speed, high-definition Image of thermal internal printer.
④Strip table unit:Place the urine strip.
⑤Serial port: RS-232C outlet with 9-core.
⑥Parallel port: Outlet with 25-core for external-printer.
⑦Power supply wire port: IEC standard three-wire for power plug.
⑧Fuse socket: FUSE.
⑨Power switch:”∣” on,”○” off.
3.2 Running Environment
Environment Requirements as below chart:
Parameters
Running Environment
Store
Environment
Temperature
0℃~40℃
0℃~40℃
Relative
Humidity
≤85%
≤85%
①
②
③
④

YD-200 Urine Analyzer User Manual
5
Barometric
Pressure
76-106Kpa
76-106Kpa
Table-board
Please put the instrument on a stable,
ventilate, clean and flat table-board and
with stable power supply, no strong
electromagnetic interference.
Note:put the instrument on a stable and flat table-board.
3.3 Functions keys
Name
Function
Menu(FN)
Start the pop-up menu.
Print(PRN)
Enabled Printing.
Feed Paper(FP)
Feed paper.
Back(Home)
Once pressed, the cursor will return to the head of the line.
UP(PgUp)
Make the display return to the upper page.
End(End)
Once pressed, the cursor will return to the end of the line.
↑ ↓ → ←
Cursor direction.
Down(PgDn)
Go forward to next page.
Esc(ESC)
Stop current option. Exist from current sub-menu, return to
upper menu. Under “start menu”, loader goes back to inside of
instrument.
Start(START)
Go to the selected sub-menu. Confirm the current correction and
start-up function.
3.4 Power on
Connect the power wire, make the power like “∣”this position, display the boot screen,
meanwhile each system begin to self-checking process, then the screen display ready
for test picture,figure3-3 after self-checking done.

YD-200 Urine Analyzer User Manual
6
#0001 4-011-01
12:00:00
VERSION 1.001
Figure 3-3
3.5 Install printing paper
3.5.1 Press the printer cover (Figure 3.3), printer cover bounce automatically, then open
the printer cover, put the new roll of paper into the paper box.
3.5.2 Insert the printing paper to make the paper top to the underside of the rubber roll,
press the FP button until the paper top is 3-4cm higher than the interface of instrument.
3.5.3 Unfold the printer paper, make the printer paper top goes through the exit of paper
box, remove the printer cover.
Note: thermal surface of the printer paper should forward to the operator while
installing paper, normally the external surface the thermal surface.
3.6 Menu Functions
Under the boot screen (Figure 3.3), press menu option (FN), another pop-up men will
be displayed, all the settings of testing, related information of hardware will be
contained in this page. The contents are as below chart:
First Display
Second Display
Third Display
Function
No. Setting
In put No.:0001
Setting sample No.
Setting
Sample
System testing
Standard strip testing
No.QC Liquid
0001
QC testing
System testing
Data Recall
Current record
the current testing result Query
Query the sequence
NO.
Input No.:0001
The sequence NO. Query
Control query
Data clear
Clear memory:YES
NO
Clear the store data

YD-200 Urine Analyzer User Manual
7
Test Mode
Fast mode
Fast mode, 120tests/hour
Slow mode
Slow continuous mode, 60tests/hour
Slow step mode
Slow mode, 60tests/hour
Set up Menu
Set clock
Set time
Set time, date, date format
Set date
Date format
Strip Select
URS-10A
URS-11A
Strip select
Status
Plus on
User can set up the Plus, External
printer, Internal printer, PC port.
External printer off
Internal printer on
PC port on
Languages
Chinese English
Languages set
Set Output
SI Conventional Select the unit of the format Plus
system
Baud Rate
1200 9600 19200
115200
Select the rate of the instrument port.
Chapter 4 Set up
4.1 Set Sequence Number
#0001 4-011-01
12:00:00
1.006
YIDA MEDICAL
Figure 4-1
In the main screen (Figure 4-1) press the key “Menu”, the screen shows as follows:
MAIN MEN
Set Seq No. Data Recall
Calibration Test Mode
QC Liquid Set Up
Figure 4-2
Press the key “Menu” return to the main screen (Figure 4-1).

YD-200 Urine Analyzer User Manual
8
Press the key “▲” and “▼” to move the cursor to the position of “Set Seq. No.”,
Press the key “START” for confirmation, and enter sequence number setup Submenu.
(Figure 4-3).
Please Input Sequence No.
Sequence No. 0001
Figure 4-3
The sequence number can be any numerical value between 0001-9999, the sequence
number will add 1 automatically after each test.
Cursor will stay at first digit position when the user goes to the submenu of “Set
Seq. No.” input the numbers by the number key. The cursor will shift to right 1 position
when inputting is finished .Setting is finished, press the “START” button to return and
vacancies will full filled with“0”.
Press the key “Menu” to discard changes and return to the upper menu (Figure 4-2).
4.2 Data recall
In main menu (Figure 4-2), press the key “Data recall” and “Start” to confirm, the
screen will show as Figure 4-4:
Current Record
Locate
QC Record
Clear Memory
Figure 4-4
User can recall the data by Current record recall and Sequence number Recall.
4.2.1 Current record
Select Current Record and press “Current record recall” and press “START”, the screen
will show current records. Press the key “DOWN” could turn record page and press the

YD-200 Urine Analyzer User Manual
9
key “▲”and“▼”to select the corresponding record as Figure 4-4:. Press the key “Print”
to print the showed record.
4.2.2 Sequence number Recall
Select the “Sequence number recall” and press the key “start” to confirm, the screen
will show as Figure 4-5:
Please Input Sequence No.
Sequence No. 0001
Figure 4-5
The cursor stops at the first number, enter the recall sequence number through the
number key. And then, press the key “START” to check the result by the sequence
number. If there is record in the instrument, the screen will display the record.
If there is no such record, the screen will display “Record Not Found!’ (Figure 4-6).
Press the “back” return to Figure 4-5:
Record Not Found#
Figure 4-6
4.2.3 Quality Control Record
Analyzer could store 50 pieces QC records. Choose “QC Record” in Figure 4-4 and
then press “start” key to confirm. Press the key Line could turn record page and press
the key “▲” and “▼” to select the corresponding record. Press the key “Print” to print
the showed record. Press “back” to return. If there is no such record, the screen will
display “Record Not Found!’
Inhaltsverzeichnis
Andere MR Messgerät Handbücher

MR
MR MESSKO EI100 Bedienungsanleitung

MR
MR MESSKO EI100 Bedienungsanleitung

MR
MR MSENSE BM Bedienungsanleitung

MR
MR MSENSE DGA 2 Bedienungsanleitung

MR
MR MSENSE BM Bedienungsanleitung

MR
MR ETOS IM Bedienungsanleitung

MR
MR TAPGUARD 260 Bedienungsanleitung

MR
MR MSENSE BM Bedienungsanleitung

MR
MR MESSKO BeTech Bedienungsanleitung

MR
MR MESSKO MFLOC 2.0 Bedienungsanleitung