(5) Upon completion of measurement, turn off the indicator.
(6) Close the cap for the refill port.
(7) Before storing, thoroughly wash the glass electrode with distilled, pure or tap water and cover the
electrode with the cap filled with water.
* For details on handling the indicator, refer to the instruction manual provided with the indicator.
Maintenance
1. Glass electrode: When the glass electrode becomes dirty, clean it as follows.
- For organic stains
Lightly wipe the electrode with gauze or cotton wool damped with a diluted neutral detergent, and
then rinse the electrode well with pure water.
- For inorganic stains
Immerse the glass electrode in an approximately 0.1N hydrochloric acid solution or a neutral detergent,
and then rinse the electrode well with pure water. Do not leave the electrode in the solution for a long time.
Note: Make sure that the neutral detergent does not come in contact with the junction. Otherwise, the
performance of the inner solution could be adversely affected if the neutral detergent penetrates inside
the sensor.
2. Reference electrode
(1) When the inner solution level falls below the position of the internal reference electrode, refill the
electrode with the inner solution. When refilling, pour an appropriate amount of the provided inner
solution through the refill port on the unit.
(2) When the inner solution becomes dirty, first remove the dirty inner solution from the holding tube,
thoroughly clean the inside of the tube with new inner solution, and then fill the electrode with new inner
solution. The inner solution should be replaced at regular intervals.
Important: The unit cannot be repaired due to its characteristics and structure. If the glass electrode is
damaged or its performance is deteriorated, replace with a new sensor probe.
Caution
For proper use of the unit, be sure to observe the following:
- The sensor is made of glass and is a precision instrument; be careful not to drop or knock it.
- If the unit is used in an environment where electrical noise is generated, the display may become unstable
or the measurement error may increase.
- Do not alter, pull forcibly, bend or bind the cord, as doing so may cause a break or malfunction in the probe.
- Never use the unit outside the measurement range, otherwise it may malfunction.
- When measuring, do not let the sensor probe come in contact with the human body.
- The glass electrode could be damaged or its service life could be shortened if it is used for measurement
samples such as organic solvents, oils, adhesives and surface-active agents.
- The pH reading may become unstable when pure water is measured.
- If the pH standard solution contacts the skin (hand), immediately wash the contaminated skin with running
water. If the solution gets into the eyes, immediately rinse with water and consult a doctor.
- Keep the pH standard solution out of reach of children. If accidentally ingested, consult a doctor immediately.
- If the unit is left in a car in the hot summer sun, it will become extremely hot and may malfunction. Do not
leave the unit in such a place.