tagarno FHD UNO Bedienungsanleitung

VERSION: 1.8 | 2019-05-08
1
Digital Microscope Solutions
CONTENTS
1. INTENDED USE 2
2. WARNINGS 2
3. TIPS 2
4. YOU HAVE RECEIVED 3
5. ASSEMBLING 4
6. CONNECTING 6
7. OPERATION 7
8. SETUP MENU 10
9. SYSTEM UPDATE 15
MANUAL | TAGARNO FHD UNO
10. LICENSE ACTIVATION 17
11. TECHNICAL SPECIFICATIONS 18
12. SYSTEM REQUIREMENTS 21
13. MAINTENANCE 22
14. WARRANTY 22
15. LICENSE AGREEMENT 22
16. ERGONOMIC RECOMMENDATIONS 22
17. DECLARATION OF CONFORMITY 23
18. COMPLIANCE STATEMENTS 24
Go to tagarno.com/productmanuals to nd manuals and tutorial videos.

2
1. Read the manual before you use the product
2. Use the product only as specied, or the protection supplied by the product can
be compromised
3. Do not position the equipment so that it is difcult to operate the disconnecting
device (appliance inlet of external power supply, equipment input connector)
4. If fluids are spilled on the product, turn the system off immediately by pulling the power
supply out of the electrical outlet
5. In case of re close to the microscope, please turn off and disconnect the system
6. Avoid subjecting the lens to sharp or hard objects
7. Please do not connect the microscope, if visible damages appear
8. Do not dismantle any parts of the microscope, except where noted in the manual
9. Never disassemble or clean internal optical surfaces
10. Use only the power supply provided from TAGARNO
11. Always turn off the system before unplugging, when possible
WARNINGS
The product is a digital magnifying system consisting of a camera unit, PCBs, mechanical
parts and a power supply. The product is intended for marketing worldwide and is designed for
manual visual inspection.
INTENDED USE
Read all safety information before you use the product.
Please pay attention when you see a warning label on the product.
This product is for indoor use only.
You must not discard this electrical/electronic product in domestic household waste.
Please dispose at your local recycling centre.
1. Avoid touching the lens glass
2. Use both hands to move the XY table
3. Use both hands to adjust the height of the product
1. INTENDED USE
2. WARNINGS
3. TIPS
12. Do not touch the light source. It gets hot when using the product
13. Do not look directly into the light source
14. Connect the power plug to a grounded power outlet
WARNING: This is a Class A product. In a domestic environment, this product may cause
radio interference, in which case the user may be required to take appropriate measures.

3
Rod, washers, screws USB memory stickManual Camera arm
USB 3.0 A-B cablePower supplyHexagon tool
4. YOU HAVE RECEIVED
OPTIONS
XPLUS FHD
control box
XKEY control box Base XY table
HDMI - HDMI cable

4
TIP!
Find the "How to Install" tutorial video here:
www.tagarno.com/installuno
Assemble base and rod with washer and screw.
Lock the camera arm and screw the washer on top of the rod.
Place the camera arm on the rod.
The complete microscope with base.
ASSEMBLING | BASE (1/2)
5. ASSEMBLING

5
Assemble XY table, block and rod with three screws.
Lock the camera arm and screw the washers on top of the rod.
Place the camera arm on the rod.
The complete microscope with XY table.
ASSEMBLING | XY TABLE (2/2)

6
5.
6.
8.
1.
2.
7.
3.
4.
6. CONNECTING
CONNECTING | BASE + XY TABLE (1/3)
Connect USB 2.0 (see options below) (1+2), control box (3),
monitor (4), ethernet (5), USB 3.0 (6), ESD (Only black models) (7)
and power supply (8) in that order.
USB 2.0 (1+2): Keyboard, mouse, USB key, USB hard drive, scanner
Connect keyboard and mouse in the USB 2.0 inputs (only for use
with some applications). Keyboard and mouse need to be HID
compliant.
You can also use the USB inputs for a USB key, USB hard drive
and/or barcode scanner. If necessary use a USB hub to get
multiple USB ports.
Connector Description Type Connect to Specication
1+2 USB 2.0 input USB 2.0 Type A Socket Keyboard
Mouse
USB key
USB hard drive
Barcode scanner
USB 2.0 Host. Output: 5V 500mA
3Control box D-SUB 9-Pin Female Control box Output: 3.3V 6A Max
Use only control boxes supplied by TAGARNO
4Output HDMI Standard Type A Monitor HDMI Out 1080p60
5Ethernet RJ-45 Ethernet wall socket Ethernet LAN, 100BASE-TX/1000BASE-T
6USB 3.0 output USB 3.0 Type B Socket Computer USB 3.0 Device. Self-powered
7ESD Snap Socket, 10mm ESD Grounding point ESD Grounding point. Only black models
8Power supply
(DC)
DC Jack Power socket Input: 12V 1.3A
Use only the power supply provided by TAGARNO:
Brand/Model: Mean Well/GSM40A12
Protection Class I
Input: 100-240V 1.0-0.5A, 50/60Hz
Overvoltage category ll
Output: 12V 3.34A Max
HDMI - HDMI cable
Live image
The microscope can be connected directly to a monitor.
Live image and image capturing through USB3.0 / 3rd part software.
See system recommended (third-party) under system requirements.
You are still able to connect monitor(s) as described above.
Live image
/ image capture
USB3.0

7
TIP!
Find the "How to Use" tutorial video here:
www.tagarno.com/useuno
Power on/off.
OPERATION | BASE + XY TABLE (1/3)
Lock the camera arm at the ring that ts the lens you are using.Adjust the height of the camera arm according to the mounted lens.
The three coloured rings on the rod mark the correct focus height for
a +3, +4 or +5 lens respectively.
Blue
Red
Black
7. OPERATION
Lens Ring colour Lens to object distance
+3 Blue 333 mm/13.1”
+4 Red 250 mm/9.8”
+5 Black 200 mm/7.9”
Lens to object distance with a +10 lens is 100 mm/3.9”.

8
XPLUS FHD Control box for TAGARNO FHD functions.
Do only connect equipment distributed by TAGARNO.
Dimensions
H: 45mm/1.8" | W: 120mm/4.7" | D: 150mm/5.9".
On screen display (OSD)
The on-screen display provides you with the relevant information
when you push a button on the control box or Setup menu.
For instance if you change the level of magnication, the white
balance, iris, etc. In this way you always know which function you are
activating and which parameters you are using.
Setup menu
Access the Setup menu by pressing the round center
button for 1 second. See separate section regarding
the Setup menu.
Snapshot function
To take a picture and transfer it to USB or your microscope
internal storage, press the image capture button until you
hear a beep. Long press on the same button will save an
image with graphics, causing the live image to freeze for a
few seconds. This button also turns Auto Snapshot on/off
when activated via the Setup menu.
When taking a picture, an information window will appear
for a few seconds showing the le name and storage
location.
Zoom
Push this button on the control box to zoom in.
Push this button on the control box to zoom out.
The magnication level will show via On Screen Display.
Auto exposure mode
Auto exposure mode prioritizes the exposure time
which is optimal for the selected video format. In good
lighting conditions the microscope will reduce the iris
instead of the exposure time to improve the depth of view.
Manual focus
Turn autofocus off by pushing both zoom buttons (plus
and minus) on the control box at the same time until you
hear a beep. You can now adjust focus manually with the
help of the zoom buttons.
Push this button on the control box to adjust focus if you
wish to focus on something very close.
Push this button on the control box to adjust focus if you
wish to focus on something that is farther away.
To turn the autofocus on again, simply push both zoom
buttons (plus and minus) at the same time until you hear
a beep. The focus values will show via On Screen Display.
Save image on USB memory stick
The image will be saved on the inserted USB memory stick.
However, if your USB memory stick is not compatible with your
microscope, a pop-up warning will instead ask you to format your
USB memory stick (for more information, see section 5 in Set-up
Menu).
When clicking OK, a second pop-up will ask you to conrm your
selection. All les on the USB memory stick will be deleted once
formatting begins. Formatting time will vary.
Press the image capture button again to take a new photo and
save it directly to your USB memory stick.
Save image on microscope
If a USB memory stick is not attached, the image is saved internally
on the microscope, provided the share le mode is turned on.
Otherwise, an error message will appear and you are not able to
save images.
To access the les on the microscope, follow descriptions regarding
le sharing in section 2 in SETUP MENU.
OPERATION | XPLUS FHD CONTROL BOX (2/3)

9
Memory Function (Preset Button)
To save a favourite setting, press the "P" button.
A Preset Manager menu appears. Here you can choose
which of the 6 presets to use/overwrite, this is done by
marking and choosing save for the given preset. To prevent
users from accidentally overwriting a saved preset, please
see section Setup Menu.
To use a preset, simply press "P" and press the center
button to choose the preset you want to recall.
The preset values will show via On Screen Display.
To rename a preset, simply choose the given preset in the
Preset Manager and press "F2". Complete the renaming by
pressing "Enter". Please note that presets can be renamed
regardless of whether the presets are locked or not.
Laser pointer (TAGARNO FHD TREND)
Press this button to turn the laser pointer on/off.
You will hear a warning when turning the laser pointer on.
NB! Never look directly into the laser pointer. The laser
pointer can only be used with lenses +3, +4 and +5.
If you are using another TAGARNO unit than TAGARNO
FHD TREND, a window on the screen will provide you with
information.
Light on/off
Press and hold this button to turn the light on/off.
When the light is turned on/off, a white balance calibration
will be performed automatically.
Application shortcut
Program an individual application shortcut by long-
pressing the button and choosing a listed application.
You are now able to quickly access this application by
short-pressing the button.
Factory default directs to the Ruler application with both
horizontal og vertical ruler activated.
Multi Function Buttons
When the Setup menu is deactivated, you are able to use
the Multi Function Buttons to control the iris, gain and
exposure time.
Press the center button to switch between auto and
manual exposure mode.
491190 - Folie til XPLUS FHD - 13-06-2017
Zoom
Push this button on the control box to zoom in.
Push this button on the control box to zoom out.
The magnication level will show via On Screen Display.
Manual Focus
Turn autofocus off by pushing both zoom buttons (plus
and minus) on the control box at the same time. You
can now adjust focus manually with the help of the
zoom buttons.
Push this button on the control box to adjust focus if you
wish to focus on something very close.
Push this button on the control box to adjust focus if you
wish to focus on something that is farther away.
To turn the autofocus on again, simply push both zoom
buttons (plus and minus) at the same time.
The focus values will show via On Screen Display.
Multi Function Buttons
Use the Multi Function Buttons to control the iris, gain
and exposure time.
Press the center button to switch between auto and
manual exposure mode. When the On Screen Display is
open, all four arrows can be used for navigating the menu
and adjustment of settings.
The window will automatically close after 3 seconds
inactivity.
XKEY Control box for TAGARNO FHD functions.
Do only connect equipment distributed by TAGARNO.
Dimensions
H: 22mm/0.87" | B: 61mm/2.40" | D: 100mm/3.94"
OPERATION | XPLUS FHD + XKEY CONTROL BOX (3/3)

10
The setup menu in your TAGARNO FHD microscope provides you
with 2 entrances to setup your microscope exactly as you like it.
Access the Setup menu by pressing the round center
button on the control box for approx. 1 second.
The 5 possibilities are:
1. Camera settings 2. General settings
3. Applications 4. Information
5. On/off
To navigate in the set-up menu, use the multi function
buttons on the control box.
When applicable, sliding the button to the right, will activate
the feature. Sliding the button left will deactivate the
functionality.
Exit setup menu
You can always exit the setup menu by pressing the center button on
the control box for 1 second or by choosing Exit in the Setup menu.
Same procedure is valid regardless of the menu option accessed.
1. CAMERA SETTINGS
1. White balance calibrate
Perform an automatic white balance adjustment in order to get
colors displayed correctly on screen. There must be a white object
present in the camera’s eld of view when the white balance is being
adjusted. A white balance calibration is made each time the camera
is turned on, to adjust to the given light conditions.
NB! If the camera’s eld of view does not contain a white object,
when the unit is turned on, the white balance will be incorrect and it
will be necessary to perform the white balance adjustment process
as described previously.
2. Optimal working distance
This setup works in 3 steps and needs to be carried out on a flat
surface. Take 3 zoom levels into consideration to perform as precise
measurements as possible. The magnication settings prior to
performance will be restored.
3. Closeup Lens
Indicate which lens is mounted on your TAGARNO microscope,
to get the right magnication levels displayed in the On Screen
Display when zooming in and out.
4. Start magnication
Choose the start magnication level you want your TAGARNO
microscope to automatically set when turning your system on.
5. Lock magnication
Lock your magnication level, so your microscope is always
working at your start magnication level and you are not able to
manually zoom in/out.
6. High contrast
This feature can be activated to visually enhance the edges of an
object.Deactivated by default.
7. Wide dynamic range
When this option is selected and auto exposure mode is active, the
camera uses individual exposure settings for dark and light areas. This
option can be useful to reduce glare when inspecting blank objects.
8. Picture ip
Select if you want to display your image on screen as placed below
the camera or flipped 180 degrees horizontally.
9. IR lter
Turn Infrared lter off when using Infrared ring light.
To use regular ring lights turn the Infrared lter on.
10. White balance calibration at LED light change
Switch on if you want an automatic white balance calibration to be
performed when LED light is changed.
Switch off if you do not want an automatic white balance calibration to
be performed when LED light is changed
8. SETUP MENU
SETUP MENU (1/5)
Inhaltsverzeichnis
Andere tagarno Mikroskop Handbücher
Beliebte Mikroskop Handbücher anderer Marken

Thermo Scientific
Thermo Scientific Nicolet iN10 Leitfaden

National
National DC3-163 Bedienungsanleitung

KERN Optics
KERN Optics OPE-1 Bedienungsanleitung

OPTIKA MICROSCOPES
OPTIKA MICROSCOPES B-150D Bedienungsanleitung

Prior
Prior Lux POL Bedienungsanleitung

Opitec
Opitec Smartscoop 115615 Bedienungsanleitung

VWR
VWR VisiScope 384 Series Bedienungsanleitung

Nikon
Nikon ECLIPSE E200 POL Bedienungsanleitung

Leica
Leica DI C800 Handbuch

ThermoFisher Scientific
ThermoFisher Scientific Continuµm Bedienungsanleitung

ThermoFisher Scientific
ThermoFisher Scientific Continuµm Bedienungsanleitung

Olympus
Olympus SZ61 Bedienungsanleitung